The 9 common sense of Rockwell hardness tester
1, Rockwell hardness elastic deformation effect on the hardness values: Rockwell  hardness up to 0.5HR effect of elastic deformation on hardness value, so we must  pay attention to the frame rigidity. The same, the machining accuracy of the  elevating screw should be higher, can by itself weight of inserting the guide  device, no obvious gap, in order to reduce the axial displacement, improving  test precision.
 2, Rockwell hardness test: must ensure that  the applied force perpendicular to the test surface. The specimen for bending  shape and other irregular shapes, must use special operation table corresponding  to the type, and select the correct test device. For cylindrical specimens, must  use the "V" type operation platform.
3. Stability test of Rockwell  hardness tester: Rockwell hardness is often adjust the measuring lever, should  maintain its stability. Available in one or two sets of good standard hardness  blocks, regularly at the same hardness block test, with long-term stability test  instrument, grasp their performance. The hardness test, standard hardness block  near the school machine application and hardness values, make the test results  more reliable.
4,  Effect of loading rate Rockwell hardness on the  hardness values: the effect of loading rate on the hardness values: when the  main load Rockwell hardness meter application acceleration from 2 seconds to 12  seconds when the effects on the hardness, low hardness value is 0.2HRC, hardness  0.4HRC, high hardness of 0.6HRC.
5,  The cone angle of the surface  Rockwell hardness: head surface Rockwell hardness value increases with the  diamond indenter cone angle. For 30N, 45N gauge, angle has great influence on  the low hardness, little influence on high hardness. The 15N scale and 30N scale  high values of hardness blocks almost no effect.
6, Rockwell hardness  surface detection: detection of sample is generally flat, if the test was  carried out on surface samples, and the curvature radius is small, the test  results should be combined with correction. The convex surface samples, the  hardness values decrease, should be amended amount plus; concave sample should  be subtracted from the correction.
7, Digital Rockwell hardness boot does  not show the reason: the power is disconnected, check whether the power line is  conducted; fuse, fuse box out attachment replacement.
8, The surface  Rockwell hardness meter installation note: open the box cover and around the  baffle at the bottom of the fuselage, demolition of 4 fixed screws, can be  placed in a vibration free hardness of dry clean and no corrosive gas and around  the room. Install the hardness of the table should be strong, and do the  diameter greater than 50 mm hole, for through screw.
 9, The adjustable level of Rockwell  hardness tester: adjustment should be the operation station installed in the  screw at the top level, then the level on the operating table, adjust, make the  hardness level within 0.2/1000.