How to detect the clean bench and biological safety cabinet qualified
 A, Clean bench
Clean bench can protect the  sample, can not protect the operators, the certification of qualified comprises  the following steps:
1, using a rotary impeller anemometer and thermal  anemometer measuring ultra clean air rate. Safety mean values: 0.4-0.5m/s, the  maximum deviation of 20%.
2, measuring the purification table filter  integrity: the use of natural wet agent detector.
3, measuring the  purification table internal intensity, noise.
B, Biological safety  cabinet
Biological safety cabinet at present and the research is closely  related to the two level of biological safety cabinets and three stage  biological safety cabinet. The two stage biological safety cabinet can be  divided into several categories, but the laboratory often is the use of 30% gas  efflux, type B2 70% gas circulation type A/B3 and 100% gas efflux. Biological  safety cabinets can not only protect the sample can also protect the operator,  so called "safe" cabinet. The detection process is as follows:
1, Into  the gas flow measurement (forward volume): by thermal anemometer or DIM air  bell.
2, To measure the downward gas flow: according to NSF49 and EN12469  standards are different, there are two ways. But it is the table to determine  the downward airflow rate using heat wind speed.
3, ULPA filter  detection: wet agent photometer, wet agent generator, pressure measuring  instrument, PAO instrument.
4, There is noise, vibration intensity,  intensity measurements.
5. Airflow image test , the smoke generator.